Metal Shards on the Reprocessed Load: Tracing Contamination Beyond the Washer-Disinfector

An investigation into metallic contamination found on reprocessed reusable medical devices demonstrates why incidents must be traced through equipment, water supplies, supporting infrastructure and maintenance history.

By admin 4 November 2026 1 min read

Why contamination investigations need to extend beyond the washer-disinfector

The discovery of visible contamination on reusable medical devices can immediately raise questions about the performance of cleaning equipment. However, washer-disinfectors do not operate in isolation. Their performance depends on the condition and quality of the water supplies, treatment systems, pipework, filters and other infrastructure supporting the reprocessing process.

This became evident when a healthcare organisation identified small metallic fragments on reusable medical devices following automated cleaning and disinfection. The fragments were initially observed on plastic orthopaedic trial devices and were subsequently found elsewhere in the reprocessing environment.

The immediate concern was clear: where were the fragments coming from, how widely had they travelled, and what controls were needed while the source was investigated?

The challenge of tracing the contamination

The presence of metallic material did not, by itself, establish its source. Possible causes included:

  • wear or deterioration within the washer-disinfectors
  • damage to instrument carriers or other moving components
  • debris introduced through the hot or cold water supplies
  • contamination within the reverse osmosis water system
  • material released during recent servicing, repairs or component failure.

A structured investigation was required to distinguish between these possibilities.

Samples and captured debris from the hot, cold and reverse osmosis water supplies were compared with the material found on the processed devices. Washer-disinfector holding tanks were opened and inspected, components of the reverse osmosis system were examined, and recent service records were reviewed for events that could have introduced contamination.

This approach was important because the location where contamination is discovered is not necessarily where it originated.

1. Following the evidence upstream

The investigation found metallic particles in the reverse osmosis water supply that closely resembled the fragments recovered from the reusable medical devices. Metallic debris was also identified within reverse osmosis connections, washer holding tanks and filtration components.

Hot and cold water supplies also contained other forms of particulate material, including material resembling silt, gasket residue and thread-sealing tape. This indicated that the organisation was dealing with more than one potential source of debris. However, the metallic particles affecting the processed devices were most closely associated with the reverse osmosis supply.

A review of the reverse osmosis system’s service history provided the critical lead. Its in-line loop heating element had failed and had been replaced approximately one month before the contamination was recognised. Inspection of the failed element showed ruptured tubes capable of releasing metallic fragments into the circulating water loop.

Importantly, the heater was positioned downstream of the system’s final filtration stage. Once released, fragments could therefore circulate through the reverse osmosis loop and reach both manual draw-off points and connected reprocessing equipment without passing through another fine filter.

The eventual location of the contamination reflected the movement of water through the system, not simply the location of the original failure.  This issue also highlighted that the system didn’t provide sufficient safeguard or control of process to detect and mitigate contamination to the water supplying the critical processes.

2. Why the issue was not immediately apparent

There can be a significant delay between an event that introduces contamination and the time the contamination is recognised.

In this instance, the heating element had failed weeks before metallic particles were reported. Once introduced, fragments could remain distributed throughout the system, including within:

  • circulating pipework
  • connecting hoses
  • washer-disinfector holding tanks
  • filters and strainers
  • equipment sumps
  • less accessible internal components.

Replacing the failed heater removed the original source, but it did not automatically remove contamination already present downstream.

The fragments were eventually identified because staff were carefully examining processed devices for an unrelated residue. This illustrates the value of vigilant visual inspection and of treating unexpected findings as potential indicators of a wider process or infrastructure issue.

3. Managing the immediate operational risk

When contamination affects essential reprocessing infrastructure, the response must balance patient safety, process integrity and operational continuity.

While the system was being investigated and remediated, additional controls were introduced. These included:

  • enhanced inspection of reusable medical devices after automated washing
  • compressed-air cleaning where appropriate to assist with detecting and removing visible fragments
  • increased cleaning and flushing of washer-disinfector tanks, sumps and filters
  • monitoring for further evidence of contamination
  • targeted filtration of incoming water supplies.

These controls did not replace correction of the underlying fault. They provided additional safeguards while residual contamination was progressively removed and confidence in the process was re-established.

4. Containing contamination already in the system

The investigation identified that some existing machine inlet strainers had screen openings too large to reliably capture fragments of the size being observed.

SVA installed fine Y-strainers at selected water inlets to intercept residual particles before they entered the reprocessing equipment. The reverse osmosis system was flushed, affected filtration components were replaced, and washer-disinfector holding tanks were inspected and cleaned.

The remediation process therefore addressed both:

  1. the originating failure, involving the damaged reverse osmosis heating element and associated system arrangement; and
  2. the downstream consequences, involving fragments already distributed through the water loop, connections and reprocessing equipment.

This distinction is important. Correcting a failed component may prevent further contamination, but additional work is usually required to demonstrate that contamination already released into the system has been adequately removed.

5. Findings beyond the immediate incident

The investigation also identified issues that were not considered the primary source of the metallic fragments but remained relevant to contamination control. These included:

  • particulate material in hot and cold water supplies
  • accumulated residue within equipment holding tanks
  • wear on washer-disinfector components and carriers
  • visible particulate matter in an endotoxin filter
  • electrical heat damage associated with the reverse osmosis heater system.

These observations reinforced the need to look beyond a single root cause. An investigation may identify one principal initiating event while also revealing secondary conditions that could affect equipment performance, water quality or future contamination risk.

What healthcare organisations can learn from this incident

Supporting services form an integral part of the reprocessing process.

The condition of the final reusable medical device depends not only on the washer-disinfector cycle. Water treatment systems, storage tanks, circulation loops, heating components, filters, pipework and maintenance activities can all affect the process outcome.

Maintenance work should be assessed for downstream consequences

A component failure or repair within a water system may introduce debris, chemicals or microorganisms into downstream infrastructure. Before work begins, and again before the system is returned to service, the organisation should consider:

  • what material could be released
  • where that material could travel
  • whether existing filtration would contain it
  • which connected equipment may be affected
  • what flushing, inspection, sampling or testing is required.

A formal permit-to-work or disruption process can help ensure these risks are assessed and communicated to the relevant clinical, engineering and reprocessing stakeholders.

Filtration must be positioned to control the identified risk

A filter can only protect equipment located downstream of it. Where a heater, pump or other component capable of shedding material is positioned after the final filtration stage, the system may remain vulnerable despite having effective upstream filtration.

Filtration arrangements should therefore be reviewed as a complete system rather than as individual components.

Holding tanks and internal water pathways require attention

Contamination may settle or accumulate in washer tanks, sumps, hoses and other low-flow areas. Maintenance programs should address the cleanliness and condition of these areas, not only calibration and mechanical operation.

Where repeated debris is identified, the cleaning method and maintenance frequency may need to be reconsidered.

Visual inspection of reprocessed loads remains an important control

Process monitoring and validation are fundamental, but neither removes the need for careful inspection of reusable medical devices. Unexpected residues, fragments, staining or surface changes can provide the first indication of a developing equipment or utility problem.

Root cause analysis should follow the process pathway

A useful investigation moves systematically through the full process:

device → equipment → holding tank → inlet connection → water service → treatment system → maintenance history

This prevents the investigation from stopping at the first visible location of the contamination and helps distinguish the point of detection from the true source.

A systems-based response

This incident was not resolved by examining the washer-disinfectors alone. It required an understanding of reprocessing practice, water treatment, mechanical systems, filtration, equipment maintenance and contamination pathways.

By comparing debris across the connected services, inspecting equipment and holding tanks, reviewing maintenance history, and tracing the reverse osmosis water pathway, the investigation identified the most probable source and supported targeted remediation.

Effective investigation requires more than correcting the obvious fault. It involves tracing the process pathway, understanding where contamination could have travelled, assessing what may remain within the system, and establishing appropriate controls before confidence in the process is restored.

Perhaps the most important lesson from this incident is simple: investigate where the evidence leads, not just where the problem first appears.

In complex reprocessing environments, finding the source often means looking beyond the washer-disinfector and understanding the complete system that supports it.

How SVA helped

SVA combined practical reprocessing experience, specialist technical knowledge and independent investigation to look beyond the immediate finding and identify the factors contributing to the issue. By considering the equipment, process and reprocessing environment together, SVA was able to provide objective evidence to support corrective action and help the facility restore confidence in its reprocessing process.

This article illustrates circumstances encountered during a specific project. Findings and outcomes are specific to the circumstances described and may not be applicable to other equipment, processes or facilities.